A multicenter post-marketing evaluation of the Elixir DESolve® Novolimus-eluting bioresorbable coronary scaffold system: First results from the DESolve PMCF study

Catheter Cardiovasc Interv. 2018 Nov 15;92(6):1021-1027. doi: 10.1002/ccd.27550. Epub 2018 Mar 6.

Abstract

Objectives: To date, experience with bioresorbable scaffolds (BRS) that elute agents other than everolimus is limited. Thus, a post-marketing clinical follow-up study was conducted to evaluate the continued safety and effectiveness of the DESolve® NOVOLIMUS™ Eluting BRS as treatment for patients with stable coronary artery disease.

Background: The DESolve BRS combines a poly-l-lactide-based backbone with a biodegradable polylactide-based polymer and Novolimus, a macrocyclic lactone mTOR inhibitor.

Methods: One hundred and two patients (mean age 62 years, 77.5% male) were enrolled at 10 European sites. Comparison of baseline and post-procedural angiographic assessment was performed, and a device-oriented composite endpoint (comprising cardiac death, target vessel myocardial infarction, and clinically driven target lesion revascularization) and rate of scaffold thrombosis at 12 months were examined.

Results: The device was successfully delivered and deployed in 98.2% (107/109) of the lesions, with two failures to cross the lesion. A total of 100 patients (109 lesions) were treated with a DESolve BRS. Post-procedural angiographic assessment indicated an in-scaffold acute gain of 1.54 ± 0.44 mm, with a reduction in % diameter stenosis from 61.00 ± 11.29 to 12.69 ± 0.44. At 12 months, the device-oriented composite endpoint had occurred in 3.0% (3/100) of patients, with 1.0% (1/100) experiencing scaffold thrombosis and myocardial infarction and 3.0% (3/100) undergoing target lesion revascularization. There were no cardiac deaths.

Conclusions: Results through 12 months indicate that the DESolve BRS is a safe and effective treatment for coronary lesions, though larger, long-term prospective studies are needed.

Keywords: bioabsorbable stent; bioresorbable devices/polymers; coronary artery disease.

Publication types

  • Multicenter Study

MeSH terms

  • Absorbable Implants*
  • Adult
  • Aged
  • Aged, 80 and over
  • Cardiovascular Agents / administration & dosage*
  • Cardiovascular Agents / adverse effects
  • Coronary Angiography
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / surgery*
  • Coronary Thrombosis / etiology
  • Female
  • Germany
  • Humans
  • Italy
  • Macrolides / administration & dosage*
  • Male
  • Middle Aged
  • Myocardial Infarction / etiology
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Product Surveillance, Postmarketing
  • Prospective Studies
  • Prosthesis Design
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • Young Adult

Substances

  • Cardiovascular Agents
  • Macrolides
  • novolimus