ISO 15189:2012 requires validation of methods used in the medical laboratory, and lists a series of performance parameters for consideration to include. Although these performance parameters are feasible for clinical chemistry analytes, application in the validation of autoimmunity tests is a challenge. Lack of gold standards or reference methods in combination with the scarcity of well-defined diagnostic samples of patients with rare diseases make validation of new assays difficult. The present manuscript describes the initiative of Dutch medical immunology laboratory specialists to combine efforts and perform multi-center validation studies of new assays in the field of autoimmunity. Validation data and reports are made available to interested Dutch laboratory specialists.
Keywords: Autoantibodies; ISO 15189; Quality; Validation.
Copyright © 2018 Elsevier B.V. All rights reserved.