Improved spatial localization in magnetic resonance spectroscopic imaging with two-dimensional PSF-Choice encoding

J Magn Reson. 2018 May:290:18-28. doi: 10.1016/j.jmr.2018.03.002. Epub 2018 Mar 3.

Abstract

Purpose: Magnetic resonance spectroscopic imaging (MRSI), under low-spatial resolution settings, often suffers signal contamination from neighboring voxels due to ringing artifacts. Spatial localization can be improved by constraining the point-spread-function (PSF). Here the effectiveness of the two-dimensional PSF-Choice technique in providing improved spatial localization for MRSI is demonstrated.

Theory and methods: The PSF-Choice technique constrains the PSF to a desired shape by manipulating the weighting of RF excitation pulse throughout phase-encode steps. Based on a Point REsolved SpectroScopy (PRESS)-type sequence, PSF-Choice encoding was implemented along two dimensions to excite a two-dimensional Gaussian profile, by replacing the usual RF excitation pulse with a train of pulses that is modified at each phase-encoding step. The method was proven mathematically, and demonstrated experimentally in phantoms containing prostate relevant metabolic compounds of choline, creatine and citrate.

Results: Using a dedicated prostate-mimicking spectroscopy phantom surrounded by oil, it was found that there is significantly less signal contamination from oil for PSF-Choice encoding compared with standard phase encoding. In particular, with standard phase encoding, there was a significant difference (p = 0.014) between ratios of Choline + Creatine to Citrate for voxels well within the phantom compared to voxels within the phantom but near the boundary with oil. The ratios in boundary voxels were also significantly different (p = 0.035) from reference values obtained using the prostate phantom with no oil present. In contrast, no significant differences were found in comparisons of these ratios when encoding with PSF-Choice.

Conclusion: The PSF-Choice scheme applied along two dimensions produces MR spectroscopic images with substantially reduced truncation artifacts and spectral contamination.

Keywords: Fourier encoding; Gibbs ringing; PSF-Choice encoding; Point spread function; Prostate; RF encoding; Spectroscopic imaging; Voxel bleeding.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Artifacts
  • Electromagnetic Fields
  • Image Enhancement / methods
  • Image Processing, Computer-Assisted
  • Lipids / chemistry
  • Magnetic Resonance Imaging / methods*
  • Normal Distribution
  • Phantoms, Imaging
  • Signal-To-Noise Ratio