Thoracic-abdominal imaging with a novel dual-layer spectral detector CT: intra-individual comparison of image quality and radiation dose with 128-row single-energy acquisition

Acta Radiol. 2018 Dec;59(12):1458-1465. doi: 10.1177/0284185118762611. Epub 2018 Mar 23.

Abstract

Background: A novel, multi-energy, dual-layer spectral detector computed tomography (SDCT) is commercially available now with the vendor's claim that it yields the same or better quality of polychromatic, conventional CT images like modern single-energy CT scanners without any radiation dose penalty.

Purpose: To intra-individually compare the quality of conventional polychromatic CT images acquired with a dual-layer spectral detector (SDCT) and the latest generation 128-row single-energy-detector (CT128) from the same manufacturer.

Material and methods: Fifty patients underwent portal-venous phase, thoracic-abdominal CT scans with the SDCT and prior CT128 imaging. The SDCT scanning protocol was adapted to yield a similar estimated dose length product (DLP) as the CT128. Patient dose optimization by automatic tube current modulation and CT image reconstruction with a state-of-the-art iterative algorithm were identical on both scanners. CT image contrast-to-noise ratio (CNR) was compared between the SDCT and CT128 in different anatomic structures. Image quality and noise were assessed independently by two readers with 5-point-Likert-scales. Volume CT dose index (CTDIvol), and DLP were recorded and normalized to 68 cm acquisition length (DLP68).

Results: The SDCT yielded higher mean CNR values of 30.0% ± 2.0% (26.4-32.5%) in all anatomic structures ( P < 0.001) and excellent scores for qualitative parameters surpassing the CT128 (all P < 0.0001) with substantial inter-rater agreement (κ ≥ 0.801). Despite adapted scan protocols the SDCT yielded lower values for CTDIvol (-10.1 ± 12.8%), DLP (-13.1 ± 13.9%), and DLP68 (-15.3 ± 16.9%) than the CT128 (all P < 0.0001).

Conclusion: The SDCT scanner yielded better CT image quality compared to the CT128 and lower radiation dose parameters.

Keywords: CT; Computed tomography; abdomen; adults; radiation safety; thorax.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Male
  • Middle Aged
  • Radiation Dosage*
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Radiography, Abdominal / methods*
  • Radiography, Thoracic / methods*
  • Reproducibility of Results
  • Tomography, X-Ray Computed / instrumentation*
  • Tomography, X-Ray Computed / methods*