The aim of this work was to assess the influence of preanalytical variables on the stability of two endogenous opioid peptides (Methionine-Enkephalin and Leucine-Enkephalin) in human plasma. For this purpose, first a sensitive LC-MS/MS analytical method was developed and validated for the simultaneous quantitative analysis of these two peptides. The methodology consisted of a simple protein precipitation step followed by UPLC separation and MRM quantitative analysis using a stable isotope labelled Methionine-Enkephalin as internal standard. The method with a limit of quantitation of 10 pg/mL showed good reproducibility with excellent accuracy and precision, and was linear up to 2000 pg/mL. An extensive evaluation of the pre-analytical stability of these peptides in human blood was carried out to ensure an adequate sample collection procedure to obtain reliable results in the analysis of clinical samples.
Keywords: Clinical samples; Enkephalins; Human plasma; LC; Leu-Enk; MS; Met-Enk; Stability.
Copyright © 2018 Elsevier Inc. All rights reserved.