Predicted longevity of contemporary cardiac implantable electronic devices: A call for industry-wide "standardized" reporting

Heart Rhythm. 2018 Dec;15(12):1756-1763. doi: 10.1016/j.hrthm.2018.07.029. Epub 2018 Jul 29.

Abstract

Background: Battery longevity is an important factor that may influence the selection of cardiac implantable electronic devices (CIEDs). However, there remains a lack of industry-wide standardized reporting of predicted CIED longevity to facilitate informed decision-making for implanting physicians and payers.

Objective: The purpose of this study was to compare the predicted longevity of current generation CIEDs using best-matched CIEDs settings to assess differences between brands and models.

Methods: Data were extracted for current model pacemakers, implantable cardioverter-defibrillators (ICDs), and cardiac resynchronization therapy-defibrillators (CRT-Ds) from product manuals and, where absent, by communication with the manufacturers. Pacemaker longevity estimations were based on standardized pacing outputs (2.5V, 0.40-ms pulse width, 500-Ω impedance) and pacing loads of 50% or 100% at 60 bpm. ICD and CRT-D longevity were estimated at 0% pacing and 15% atrial plus 100% biventricular pacing, with essential capacitor reforms and zero clinical shocks.

Results: Mean maximum predicted longevity of single- and dual-chamber pacemakers was 12.0 ± 2.1 and 9.8 ± 1.9 years, respectively. Use of advanced features such as remote monitoring, prearrhythmia electrogram storage, and rate response can result in ∼1.4 years of reduction in longevity. Mean maximum predicted longevity of ICDs and CRT-Ds was 12.4 ± 3.0 and 8.8 ± 2.1 years, respectively. Of note, there were significant variations in predicted CIED longevity according to device manufacturers, with up to 44%, 42%, and 44% difference for pacemakers, ICDs, and CRT-Ds, respectively.

Conclusion: Contemporary CIEDs demonstrate highly variable predicted longevity according to device manufacturers. This may impact on health care costs and long-term clinical outcomes.

Keywords: Battery longevity; Cardiac implantable electronic device; Cardiac resynchronization therapy; Implantable cardioverter-defibrillator; Pacemaker.