For use with intact perfused organs a spectrophotometer system has been developed, both for dual-wavelength absorption measurements and for spectral scanning. A monochromator is used for illumination and scanning in the visible and near infrared. Optic fibres conduct light to the specimen under examination and from the specimen to a detecting photomultiplier. System control is exercised by a microcomputer, which also processes the collected data. The performance of the system on isolated perfused rat heart is demonstrated, in the spectral scanning mode and in the dual-wavelength mode, by studying simultaneously the kinetics of cytochrome aa3, and myoglobin oxidation-reduction.