Avoiding inappropriate therapy of single-lead implantable cardioverter-defibrillator by using atrial-sensing electrodes

J Cardiovasc Electrophysiol. 2018 Dec;29(12):1682-1689. doi: 10.1111/jce.13736. Epub 2018 Oct 5.

Abstract

Introduction: The single-chamber implantable cardioverter-defibrillator (ICD) can be associated with more frequent inappropriate therapies compared with dual-chamber ICDs, when they are accompanied by a simpler implantation procedure. The aim of this study was to investigate whether the use of a single-lead ICD system with atrial-sensing electrodes results in a reduction of inappropriate ICD therapy.

Methods and results: The study population consisted of 212 consecutive patients, who underwent primary prophylactic single-lead ICD implantation at our institute. A ventricular lead with atrial-sensing electrodes was implanted in 77 patients (36%; Group-VDD) and a ventricular lead without atrial-sensing electrodes was implanted in 135 patients (64%; Group-VVI). Procedural and follow-up data were collected in a prospective registry. A higher prevalence of atrial fibrillation was present in Group-VDD. There were no other significant differences in patient baseline characteristics (age, sex, and other comorbidities) or follow-up period between the two groups. The operative parameters including fluoroscopic burden showed no significant differences between Group-VDD and Group-VVI. During a mean follow-up period of 697 ± 392 days, 26 patients (12%) experienced appropriate ICD therapies and 13 patients (6%) suffered inappropriate ICD therapies. The incidence of inappropriate ICD therapies in Group-VDD was significantly lower as compared to that of Group-VVI (1/77 [1%] vs 12/135 [9%]; log-rank, P = 0.028). The incidence of appropriate ICD therapies and the occurrence of device-related complications showed no significant difference between the two groups.

Conclusion: Single-lead ICD with atrial-sensing electrodes shows a lower incidence of inappropriate ICD therapy compared with the absence of atrial-sensing electrodes, without additional operative burden or increased complications.

Keywords: VDD-ICD; implantable cardioverter-defibrillator; inappropriate shock; single-lead ICD.

MeSH terms

  • Aged
  • Atrial Fibrillation / physiopathology*
  • Atrial Fibrillation / surgery*
  • Defibrillators, Implantable* / standards
  • Electrocardiography / instrumentation
  • Electrocardiography / standards
  • Electrodes, Implanted* / standards
  • Equipment Design / instrumentation*
  • Equipment Design / standards
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Registries / standards