Ex vivo assessment and in vivo validation of non-invasive stent monitoring techniques based on microwave spectrometry

Sci Rep. 2018 Oct 4;8(1):14808. doi: 10.1038/s41598-018-33254-9.

Abstract

Some conditions are well known to be directly associated with stent failure, including in-stent re-occlusion and stent fracture. Currently, identification of these high-risk conditions requires invasive and complex procedures. This study aims to assess microwave spectrometry (MWS) for monitoring stents non-invasively. Preliminary ex vivo data are presented to move to in vivo validation. Fifteen mice were assigned to receive subcutaneous stent implantations (n = 10) or sham operations (n = 5). MWS measurements were carried out at 0, 2, 4, 7, 14, 22, and 29 days of follow-up. Additionally, 5 stented animals were summited to micro-CT analyses at the same time points. At 29 days, 3 animals were included into a stent fracture subgroup and underwent a last MWS and micro-CT analysis. MWS was able to identify stent position and in-stent stenosis over time, also discerning significant differences from baseline measures (P < 0.001). Moreover, MWS identified fractured vs. non-fractured stents in vivo. Taken together, MWS emerges as a non-invasive, non-ionizing alternative for stent monitoring. MWS analysis clearly distinguished between in-stent stenosis and stent fracture phenomena.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Equipment Failure Analysis / methods*
  • Mice
  • Microwaves*
  • Spectrum Analysis / methods*
  • Stents*