Purpose: At introduction in 2014, dose calculation for the first MLC on a robotic SRS/SBRT platform was limited to a correction-based Finite-Size Pencil Beam (FSPB) algorithm. We report on the dosimetric accuracy of a novel Monte Carlo (MC) dose calculation algorithm for this MLC, included in the Precision™ treatment planning system.
Methods: A phantom was built of one slab (5.0 cm) of lung-equivalent material (0.09…0.29 g/cc) enclosed by 3.5 cm (above) and 5 cm (below) slabs of solid water (1.045 g/cc). This was irradiated using rectangular (15.4 × 15.4 mm2 to 53.8 × 53.7 mm2) and two irregular MLC-fields. Radiochromic film (EBT3) was positioned perpendicular to the slabs and parallel to the beam. Calculated dose distributions were compared to film measurements using line scans and 2D gamma analysis.
Results: Measured and MC calculated percent depth dose curves showed a characteristic dose drop within the low-density region, which was not correctly reproduced by FSPB. Superior average gamma pass rates (2%/1 mm) were found for MC (91.2 ± 1.5%) compared to FSPB (55.4 ± 2.7%). However, MC calculations exhibited localized anomalies at mass density transitions around 0.15 g/cc, which were traced to a simplification in electron transport. Absence of these anomalies was confirmed in a modified build of the MC engine, which increased gamma pass rates to 96.6 ± 1.2%.
Conclusions: The novel MC algorithm greatly improves dosimetric accuracy in heterogeneous tissue, potentially expanding the clinical use of robotic radiosurgery with MLC. In-depth, independent validation is paramount to identify and reduce the residual uncertainties in any software solution.
Keywords: Inhomogeneous/heterogeneous tissues; MLC; Monte Carlo dose calculation; Robotic radiosurgery.