Midterm Single-Center Results of Endovascular Aneurysm Repair With Additional EndoAnchors

J Endovasc Ther. 2019 Feb;26(1):90-100. doi: 10.1177/1526602818816099. Epub 2018 Dec 4.

Abstract

Purpose: To review midterm clinical outcomes of EndoAnchor placement during or after endovascular aneurysm repair (EVAR) or chimney EVAR (ch-EVAR).

Materials and methods: A retrospective analysis was conducted of 51 consecutive patients [median age 75 years; 38 men] who underwent EVAR/ch-EVAR with EndoAnchor placement between June 2010 and December 2016 to prevent seal failures (31, 61%) or to treat type Ia endoleak and/or migration (20, 39%). Median aortic neck diameter was 27.7 mm and median neck length was 9.0 mm. Thirty-three (65%) had a conical neck; 48 (94%) had at least 1 hostile neck characteristic. Thirty-two (63%) patients had severe comorbidities (ASA score ⩾III). Eight patients had a single ch-EVAR procedure. Baseline patient characteristics, anatomic variables, procedure details, early and late complications, reinterventions, and aneurysm-related and all-cause mortality rates were recorded. Follow-up imaging was performed with computed tomography angiography (CTA) or duplex ultrasonography.

Results: Median procedure time was 100 minutes; a median of 6 EndoAnchors were implanted. There were 10 (10%) residual type Ia endoleaks at the end of the procedure; 9 had resolved by the first postoperative CTA. One residual and 2 new type Ia endoleaks were identified at the first postoperative imaging. Median follow-up for the entire cohort was 24.0 months, during which 3 new type Ia endoleaks were identified. Five of the 6 type Ia endoleaks were treated, 1 resolved spontaneously. There was 1 endograft limb occlusion without clinical consequences, 1 chimney graft occlusion without possibilities for a reintervention, 1 rupture after type IV endoleak (a Nellix device was successfully deployed within the main device), and 1 complete graft explantation for infection. There was no new-onset hemodialysis. Kaplan-Meier estimates of freedom from type Ia endoleak, proximal neck-related reinterventions, and aneurysm-related mortality at 2 years were 87.3%, 92.2%, and 94.0%, respectively.

Conclusion: EndoAnchors are helpful in the endovascular treatment of unfavorable proximal aortic necks, with fair midterm results.

Keywords: abdominal aortic aneurysm; chimney grafts; complications; endoanchor; endograft; endoleak; endovascular aneurysm repair; hostile neck; reintervention; stent-graft.

Publication types

  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / mortality
  • Aortic Aneurysm, Abdominal / surgery*
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis Implantation / mortality
  • Blood Vessel Prosthesis*
  • Endoleak / etiology
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation*
  • Endovascular Procedures / mortality
  • Female
  • Humans
  • Male
  • Middle Aged
  • Netherlands
  • Progression-Free Survival
  • Retrospective Studies
  • Risk Factors
  • Stents*
  • Time Factors