Purpose: This study aimed to assess the effectiveness of multiple dose-volume specifications in minimizing interinstitutional, target-prescribed, dose variations for spine stereotactic body radiation therapy (SBRT).
Methods and materials: Seven institutions with a total of 10 treatment apparatuses participated in this study. SBRT plans for 3 representative spinal metastases were generated using 2 different protocols (Protocols 1 and 2) for target dose. While using just 2 target dose objectives (doses delivered to 95% and maximum point dose) in Protocol 1, 3 target dose constraints (doses delivered to 95% and 50% and maximum point dose) were defined in Protocol 2 with the intent to decrease target dose variation. A dose-volume histogram analysis was performed for the evaluated planning target volume (PTVevl) and critical neural structures such as the spinal cord and cauda equina.
Results: Doses to the organs at risk were all maintained at the maximal tolerance in both protocols; however, the interinstitutional variation of the PTVevl dose-volume histograms was significantly decreased with Protocol 2. Furthermore, the mean PTVevl covered by the prescription dose was increased from 73.0% in Protocol 1 to 85.8% in Protocol 2. There were no differences in the mean values of the nearly maximum dose of the critical neural structures between 2 protocols.
Conclusions: In spine SBRT with the emphasis on preservation of critical neural structures, the target prescribed dose should be defined by using multiple dose-volume objectives to minimize user and apparatus-dependent dose variabilities for the spinal metastases that are adjacent to the critical neural structures.
Copyright © 2018 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.