A novel two-dimensional liquid chromatography - Mass spectrometry method for direct drug impurity identification from HPLC eluent containing ion-pairing reagent in mobile phases

Anal Chim Acta. 2019 Feb 21:1049:105-114. doi: 10.1016/j.aca.2018.10.031. Epub 2018 Oct 23.

Abstract

In this study, a novel two dimensional liquid chromatography - mass spectrometry (2D-LC-MS) method with use of a weak anion exchange column between the 1st DLC RP column and the 2nd DLC RP column (RP1-WAX-RP2) was developed and applied to identify drug impurities from MS incompatible mobile phases containing sodium 1-octanesulfonate and non-volatile buffer. The 1st DLC conditions follow exactly the original standard HPLC method recorded in Chinese Pharmacopeia (ChP), European Pharmacopeia (EP) or US Pharmacopeia (USP). An impurity fraction was collected with a built-in sample loop (100 μL) and transferred to the WAX column where 1-octanesulfonate and phosphate were trapped and removed. While, the impurity and other cations were eluted to the 2nd D column (RP2) for separation and identification by connected IT-TOF MS. Methods were programmed and applied to identify impurities in two generic drugs, sulpiride (hydrophilic drug with logP 0.57) and dobutamine (hydrophobic drug with logP 3.6). The results indicate that the methods based on RP1-WAX-RP2 column configuration offer a feasible solution for direct impurity identification in generic drug product or API without needs of off-line desalting from the MS incompatible mobile phases containing ion-pairing reagent and non-volatile buffer.

Keywords: Generic drug; Impurity; Ion-pairing reagent; Mass spectrometry; Non-volatile; Two dimensional liquid chromatography.

MeSH terms

  • Chromatography, High Pressure Liquid
  • Chromatography, Ion Exchange / methods*
  • Chromatography, Reverse-Phase / methods
  • Dobutamine / analysis*
  • Drug Contamination*
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Sulpiride / analysis*

Substances

  • Dobutamine
  • Sulpiride