Clinical experience with two-point mDixon turbo spin echo as an alternative to conventional turbo spin echo for magnetic resonance imaging of the pediatric knee

Pediatr Radiol. 2019 May;49(6):791-800. doi: 10.1007/s00247-019-04349-w. Epub 2019 Feb 6.

Abstract

Background: Two-point modified Dixon (mDixon) turbo spin-echo (TSE) sequence provides an efficient, robust method of fat suppression. In one mDixon acquisition, four image types can be generated: water-only, fat-only, in-phase and opposed-phase images.

Objective: To determine whether PD mDixon TSE water-only and, by proxy, PD in-phase images generated by one acquisition can replace two conventional PD TSE sequences with and without fat suppression in routine clinical MR examination of the knee.

Materials and methods: This is a retrospective study of 50 consecutive pediatric knee MR examinations. PD mDixon TSE water-only and PD fat-saturated TSE sequences (acquired in the sagittal plane with identical spatial resolution) were reviewed independently by two pediatric radiologists for homogeneity of fat suppression and detection of intra-articular pathology. Thirteen of the 50 patients underwent arthroscopy, and we used the arthroscopic results as a reference standard for the proton-density fat-saturated and proton-density mDixon results. We used the Kruskal-Wallis rank test to assess difference in fat suppression between the proton-density mDixon and proton-density fat-saturated techniques. We used kappa statistics to compare the agreement of detection of intra-articular pathology between readers and techniques. We also calculated sensitivity, specificity and accuracy between arthroscopy and MR interpretations.

Results: Proton-density mDixon water-only imaging showed significant improvement with the fat suppression compared with proton-density fat-saturated sequence (P=0.02). Each observer demonstrated near-perfect agreement between both techniques for detecting meniscal and ligamentous pathology and fair to substantial agreement for bone contusions, and chondral and osteochondral lesions.

Conclusion: Two-point mDixon water-only imaging can replace conventional proton-density fat-saturated sequence. When same-plane proton-density fat-saturated and non-fat-saturated sequences are required, proton-density water-only and proton-density in-phase image types acquired in the same acquisition shorten the overall examination time while maintaining excellent intra-articular lesion conspicuity.

Keywords: Adolescents; Children; Fat suppression; Knee; Magnetic resonance imaging; Modified Dixon technique.