Dynamic Modulated Brachytherapy (DMBT) Balloon Applicator for Accelerated Partial Breast Irradiation

Int J Radiat Oncol Biol Phys. 2019 Jul 15;104(4):953-961. doi: 10.1016/j.ijrobp.2019.03.026. Epub 2019 Mar 22.

Abstract

Purpose: To propose a novel high-dose-rate brachytherapy applicator for balloon-based dynamic modulated brachytherapy (DMBT) for accelerated partial breast irradiation (APBI) and to demonstrate its dosimetric advantage compared to the widely used Contura applicator.

Methods and materials: The DMBT balloon device consists of a fixed central channel enabling real-time, in vivo dosimetry and an outer motion-dynamic, adjustable-radius channel capable of moving to any angular position within the balloon. This design allows placement of dwell positions anywhere within the balloon volume, guaranteeing optimal placement and generation of the applicator and treatment plan, respectively. Thirteen clinical treatment plans for patients with early-stage breast cancer receiving APBI after lumpectomy using Contura were retrospectively obtained under institutional review board approval. New treatment plans were created by replacing the Contura with the DMBT device. DMBT plans were limited to 4 angular positions and an outer channel radius of 1.5 cm. The new plans were optimized to limit dose to ribs and skin while maintaining target coverage similar to that of the clinical plan.

Results: Similar target coverage was obtained for the DMBT plans compared with clinical Contura plans. Across all patients the mean (standard deviation) reductions in D0.1 cc to the ribs and skin were 6.70% (6.28%) and 5.13% (6.54%), respectively. A threshold separation distance between the balloon surface and the organ at risk (OAR), below which dosimetric changes of greater than 5% were obtained, was observed to be 12 mm for ribs and skin. When both OARs were far from the balloon, DMBT plans were of similar quality to Contura plans, as expected.

Conclusions: This study demonstrates the superior ability of the APBI DMBT applicator to spare OARs while achieving target coverage comparable to current treatment plans, especially when in close proximity. The DMBT balloon may enable new modes of dynamic high-dose-rate treatment delivery and allow for ultrahypofractionated dose regimens to be safely used.

MeSH terms

  • Brachytherapy / instrumentation*
  • Breast Neoplasms / radiotherapy*
  • Breast Neoplasms / surgery
  • Equipment Design
  • Female
  • Humans
  • Organ Sparing Treatments / instrumentation
  • Organ Sparing Treatments / methods
  • Organs at Risk / radiation effects*
  • Postoperative Period
  • Radiation Injuries / prevention & control*
  • Retrospective Studies