Continuous noninvasive hemoglobin monitoring estimates timing for detecting anemia better than clinicians: a randomized controlled trial

BMC Anesthesiol. 2019 May 17;19(1):80. doi: 10.1186/s12871-019-0755-1.

Abstract

Background: Hemoglobin measurement is important for transfusion decision-making. Pulse CO-Oximetry provides real-time continuous hemoglobin (SpHb) monitoring. The triage role of SpHb trends based on hemoglobin measurements was investigated.

Methods: In this diagnostic randomized controlled trial, 69 patients undergoing spine or cytoreductive surgery were randomly enrolled into SpHb-monitoring and standard-care groups. Diagnostic blood samples were drawn for CO-oximetry Hb (CoOxHb) when the SpHb decreased by 1 g/dl or at the clinician's discretion in the standard-care group. The positive predictive value (PPV) was defined as the ability to detect a decrease in CoOxHb > 1 g/dl or a CoOxHb < 10 g/dl; the PPVs were compared using Fisher's exact test. The SpHb and trend accuracies were calculated. The transfusion units and postoperative hemoglobin levels were compared.

Results: The PPV of a decrease in CoOxHb > 1 g/dl was 93.3% in the SpHb group vs 54.5% without SpHb monitoring (p = 0.002). The PPV of CoOxHb < 10 g/dl was 86.7% vs. 50.0% for these groups (p = 0.015). The CoOxHb was never < 7 g/dl with SpHb monitoring. Sixty SpHb-CoOxHb data pairs and 28 delta pairs (ΔSpHb-ΔCoOxHb) were collected. The bias, precision and limits of agreement were - 0.29, 1.03 and - 2.30 to 1.72 g/dl, respectively. When ΔSpHb and ΔCoOxHb were > 1 g/dl, the concordance rate for changes in hemoglobin reached 100%. The delta pairs revealed a positive correlation [ΔSpHb = 0.49 * ΔCoOxHb - 0.13; r = 0.69, 95% confidence interval (0.53, 0.82)]. No significant differences were found in the transfusion volume or postoperative anemia state.

Conclusions: The SpHb trend tracked changes in hemoglobin satisfactorily during surgery and more accurately estimated the appropriate timing for invasive hemoglobin measurements than the clinicians.

Trial registration: ChiCTR1800016290 (Prospective registered). Initial registration date was 24/05/2018.

Keywords: Anemia; Hemoglobin; Monitoring; Pulse CO-oximetry; Surgery; Trends.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Anemia / blood*
  • Anemia / diagnosis*
  • Erythrocyte Transfusion / methods
  • Female
  • Hemoglobins / analysis
  • Hemoglobins / metabolism*
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Intraoperative / methods*
  • Oximetry / methods
  • Time Factors

Substances

  • Hemoglobins

Associated data

  • ChiCTR/ChiCTR1800016290