Measurement of spleen fat on MRI-proton density fat fraction arises from reconstruction of noise

Abdom Radiol (NY). 2019 Oct;44(10):3295-3303. doi: 10.1007/s00261-019-02079-z.

Abstract

Purpose: This study compares splenic proton density fat fraction (PDFF) measured using confounder-corrected chemical shift-encoded (CSE)-MRI to magnetic resonance spectroscopy (MRS) in human patients at 3T.

Methods: This was a prospectively designed ancillary study to various previously described single-center studies performed in adults and children with known or suspected nonalcoholic fatty liver disease. Patients underwent magnitude-based MRI (MRI-M), complex-based MRI (MRI-C), high signal-to-noise variants (Hi-SNR MRI-M and Hi-SNR MRI-C), and MRS at 3T for spleen PDFF estimation. PDFF from CSE-MRI methods were compared to MRS-PDFF using Wilcoxon signed-rank tests. Demographics were summarized descriptively. Spearman's rank correlations were computed pairwise between CSE-MRI methods. Individual patient measurements were plotted for qualitative assessment. A significance level of 0.05 was used.

Results: Forty-seven patients (20 female, 27 male) including 12 adults (median 55 years old) and 35 children (median 12 years old). Median PDFF estimated by MRS, MRI-M, Hi-SNR MRI-M, MRI-C, and Hi-SNR MRI-C was 1.0, 2.3, 1.9, 2.2, and 2.0%. The four CSE-MRI methods estimated statistically significant higher spleen PDFF values compared to MRS (p < 0.0001 for all). Pairwise associations in spleen PDFF values measured by different CSE-MRI methods were weak, with the highest Spearman's rank correlations being 0.295 between MRI-M and Hi-SNR MRI-M; none were significant after correction for multiple comparisons. No qualitative relationship was observed between PDFF measurements among the various methods.

Conclusion: Overestimation of PDFF by CSE-MRI compared to MRS and poor agreement between related CSE-MRI methods suggest that non-zero PDFF values in human spleen are artifactual.

Keywords: Artifactual; CSE-MRI; Fat quantification; Spectroscopy; Spleen.

Publication types

  • Case Reports
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adipose Tissue / diagnostic imaging*
  • Adolescent
  • Adult
  • Algorithms
  • Child
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted
  • Imaging, Three-Dimensional
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Spectroscopy / methods*
  • Male
  • Middle Aged
  • Prospective Studies
  • Signal-To-Noise Ratio
  • Spleen / diagnostic imaging*