Artifacts in pulse transit time measurements using standard patient monitoring equipment

PLoS One. 2019 Jun 21;14(6):e0218784. doi: 10.1371/journal.pone.0218784. eCollection 2019.

Abstract

Objective: Pulse transit time (PTT) refers to the time it takes a pulse wave to travel between two arterial sites. PTT can be estimated, amongst others, using the electrocardiogram (ECG) and photoplethysmogram (PPG). Because we observed a sawtooth artifact in the PTT while using standard patient monitoring equipment for ECG and PPG, we explored the reasons for this artifact.

Methods: PPG and ECG were simulated at a heartrate of both 100 and 160 beats per minute while using a Masimo PPG post-processing module and a Philips patient monitor setup at the neonatal intensive care unit. Two different post-processing modules were used. PTT was defined as the difference between the R-peak in the ECG and the point of 50% increase in the PPG.

Results: A sawtooth artifact was seen in all simulations. Both length (59.2 to 72.4 s) and amplitude (30.8 to 36.0 ms) of the sawtooth were dependent on the post-processing module used. Furthermore, the absolute PTT value differed up to 250 ms depending on post-processing module and heart rate. The sawtooth occurred because the PPG wave continuously showed a minimal prolongation during the length of the sawtooth, followed by a sudden shortening. Both artifacts were generated in the post-processing module containing Masimo algorithms.

Conclusion: Post-processing of the PPG signal in the Masimo module of the Philips patient monitor introduces a sawtooth in PPG and derived PTT. This sawtooth, together with a large module-dependent absolute difference in PTT, renders the thus-derived PTT insufficient for clinical purposes.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Algorithms
  • Artifacts*
  • Blood Pressure / physiology
  • Blood Pressure Determination / instrumentation
  • Blood Pressure Determination / standards
  • Computer Simulation
  • Electrocardiography / instrumentation*
  • Electrocardiography / standards
  • Heart Rate / physiology
  • Humans
  • Monitoring, Physiologic* / instrumentation
  • Monitoring, Physiologic* / methods
  • Monitoring, Physiologic* / standards
  • Photoplethysmography / instrumentation*
  • Photoplethysmography / standards
  • Pulse Wave Analysis* / instrumentation
  • Pulse Wave Analysis* / methods
  • Pulse Wave Analysis* / standards
  • Reference Standards
  • Signal Processing, Computer-Assisted / instrumentation

Grants and funding

The author(s) received no specific funding for this work.