Head-to-head comparison of image quality between brain 18F-FDG images recorded with a fully digital versus a last-generation analog PET camera

EJNMMI Res. 2019 Jul 12;9(1):61. doi: 10.1186/s13550-019-0526-5.

Abstract

Background: The quality of phantom images was previously shown to be higher on digital (Vereos Philips®) compared to analog PET (Ingenuity Philips®) cameras. This study aimed to determine the extent to which this difference still remains significant on normal brain 18F-FDG PET images.

Methods: Relative noise and contrast as well as border sharpness (a spatial resolution index) of central (striata) and peripheral (occiput) gray-matter structures were compared between 10 sets of normal brain 18F-FDG PET images recorded and reconstructed on digital and analog last-generation PET cameras, together with a subjective visual analysis of image quality provided by experienced physicians.

Results: Compared with analog PET, digital PET provided marked improvements in image quality parameters. The median relative noise was decreased (- 22%), while gray/white-matter contrast was increased (+ 27%/+ 41% for central/peripheral gray-matter structures), with these results being consistent with visual analysis. In addition, a clear enhancement in image sharpness was further documented for digital PET owing to the possible use of a 1-mm3 voxel size (+ 24%/+ 21%).

Conclusions: On normal brain 18F-FDG images and compared with a last-generation analog PET, the fully digital PET camera offers marked improvements in image noise and contrast, as well as significant potential for further enhancing spatial resolution.

Keywords: Brain 18F-FDG; Contrast; Digital PET; Image quality; Noise; Spatial resolution.