Implementation of Single-Tab Electrodes for Bioimpedance Spectroscopy Measures

Lymphat Res Biol. 2020 Jun;18(3):277-283. doi: 10.1089/lrb.2019.0035. Epub 2019 Sep 19.

Abstract

Background: Bioimpedance spectroscopy (BIS) demonstrates proficiency in early identification of breast cancer treatment-related lymphedema (BCRL) development. Dual-tab electrodes were designed for consistent and easy electrode placement, however, single-tab electrodes positioned to mimic dual-tab placement on the body may make BIS technology more accessible in community hospitals and outpatient settings. The purpose of this study is to evaluate use of single-tab electrodes for BIS measurements and assess whether single-tab electrodes provide consistent BIS measurements in controls and patients with BCRL. Methods and Results: Upper limb BIS ratios were obtained using ImpediMed L-Dex® U400 in controls (n = 13; age = 23-75 years; 9 repeated measurements) using dual-tab and single-tab electrodes. BCRL patients (n = 17; Stage = 1.65 ± 0.49; number nodes removed = 16.3 ± 7.7; age = 50.9 ± 10.6 years; age range = 33-77 years) and healthy controls (n = 19) were evaluated to determine if single-tab electrodes provided discriminatory capacity for detecting BCRL. Intraclass correlation coefficients (ICC), linear mixed-effects models, Wilcoxon rank-sum tests, and linear regression with two-sided p-values <0.05 required for significance were applied. Single-tab electrodes were found to be statistically interchangeable with dual-tab electrodes (ICC = 0.966; 95% confidence interval = 0.937-0.982). No evidence of differences in single-tab versus dual-tab measurements were found for L-Dex ratios (p = 0.74) from the linear mixed-effects model. Repeated trials involving reuse of the same electrodes revealed a trend toward increases in L-Dex ratio for both styles of electrodes. Single-tab electrodes were significant (p < 0.0001) for discriminating between BCRL and control subjects. Conclusion: Findings expand upon the potential use of BIS in clinic and research settings and suggest that readily available single-tab electrodes provide similar results as dual-tab electrodes for BIS measurements.

Keywords: L-Dex; bioimpedance spectroscopy; breast cancer treatment-related lymphedema; electrodes; lymphedema; physical therapy.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Breast Cancer Lymphedema* / diagnosis
  • Breast Neoplasms* / complications
  • Case-Control Studies
  • Electric Impedance*
  • Electrodes*
  • Female
  • Humans
  • Middle Aged
  • Spectrum Analysis
  • Upper Extremity
  • Young Adult