It is common practice to utilize an internal standard (IS) to minimize variance in bioanalytical assays employing liquid chromatography coupled to mass spectrometry. For assays to be deployed in regulated drug development studies, ensuring the IS will compensate for differences in recovery, liquid handling and ionization efficiency should be determined early in the method development process. In this perspective article, we outline key considerations when selecting an IS and propose experiments to perform within the method development phase to demonstrate suitability of the IS within the assay prior to validation. Finally, a series of case studies will be presented, which illustrate analytical challenges related to internal standardization that we have observed in our laboratory.
Keywords: internal standard; liquid chromatography mass spectrometry; method development; method establishment.