Correcting Artifacts in Single Molecule Localization Microscopy Analysis Arising from Pixel Quantum Efficiency Differences in sCMOS Cameras

Sci Rep. 2019 Dec 2;9(1):18058. doi: 10.1038/s41598-019-53698-x.

Abstract

Optimal analysis of single molecule localization microscopy (SMLM) data acquired with a scientific Complementary Metal-Oxide-Semiconductor (sCMOS) camera relies on statistical compensation for its pixel-dependent gain, offset and readout noise. In this work we show that it is also necessary to compensate for differences in the relative quantum efficiency (RQE) of each pixel. We found differences in RQE on the order of 4% in our tested sCMOS sensors. These differences were large enough to have a noticeable effect on analysis algorithm results, as seen both in simulations and biological imaging data. We discuss how the RQE differences manifest themselves in the analysis results and present the modifications to the Poisson maximum likelihood estimation (MLE) sCMOS analysis algorithm that are needed to correct for the RQE differences.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Algorithms
  • Animals
  • Artifacts*
  • Calibration
  • Equipment Design
  • Image Processing, Computer-Assisted / methods*
  • Mice
  • Microscopy, Fluorescence / instrumentation
  • Microscopy, Fluorescence / standards
  • Poisson Distribution
  • Quantum Dots / standards
  • Semiconductors / standards
  • Single Molecule Imaging / instrumentation*
  • Single Molecule Imaging / standards
  • Thalamus / diagnostic imaging