Dosimetric assessment of patient-specific breath-hold reproducibility on liver motion for SBRT planning

J Appl Clin Med Phys. 2020 Jul;21(7):77-83. doi: 10.1002/acm2.12887. Epub 2020 Apr 26.

Abstract

Purpose: To investigate the impact of breath-hold reproducibility on liver motion using a respiratory motion management device.

Methods: Forty-four patients with hepatic tumors, treated with SBRT with breath-hold, were randomly selected for this study. All patients underwent three consecutive computed tomography (CT) scans using active breath-hold coordinator (ABC) with three repeated single breath-hold during simulation. The three CT scans were labeled as ABC1-CT, ABC2-CT, and ABC3-CT. Displacements of centroids of the entire livers among the three ABC-CTs were measured as a surrogate for intrafractional motion. For each patient, two different treatment plans were prepared: (a) a clinical plan using a 5-mm expansion of an ITV that encompassed all three GTVs from each of the three ABC-CTs, and (b) a research plan using a 5-mm expansion of the GTV from only ABC1-CT to create PTV. The clinical plan acceptance criteria were that 95% of the PTV and 99% of the GTV received 100% of the prescription dose. Dosimetric endpoints were analyzed and compared for the two plans.

Results: All shifts in the medial-lateral direction (range: -3.9 to 2.0 mm) were within 5 mm while 7% of shifts in the anterior-posterior direction (range: -10.5 to 16.7 mm) and 11% of shifts in the superior-inferior direction (range: -17.0 to 8.7 mm) exceeded 5 mm. Six patients (14%) had an intrafraction motion greater than 5 mm in any direction. For these six patients, if a plan was created based on a PTV from a single CT (ex. ABC1-CT), 5 of 12 GTVs captured from other ABC-CTs would fail to meet the clinical acceptance criteria due to poor breath-hold reproducibility.

Conclusions: Non-negligible intrafractional motion occurs in patients with poor breath-hold reproducibility. To identify this subgroup of patients, acquiring three CTs with active breath-hold during simulation is a feasible practical method.

Keywords: active breathing control; intrafractional motion; liver SBRT; target coverage; treatment plan.