Aims: Sacral neuromodulation (SNM) is a guideline-recommended treatment for voiding dysfunction including urgency, urge incontinence, and nonobstructive retention as well as fecal incontinence. The Axonics® System is a miniaturized, rechargeable SNM system designed to provide therapy for at least 15 years, which is expected to significantly reduce revision surgeries as it will not require replacement as frequently as the non-rechargeable SNM system. The ARTISAN-SNM study is a pivotal study designed to treat patients with urinary urgency incontinence (UUI). Clinical results at 1-year are presented.
Methods: A total of 129 eligible UUI patients were treated. All participants were implanted with a quadripolar tined lead and neurostimulator in a single procedure. Efficacy data were collected using a 3-day bladder diary, a validated quality of life questionnaire (ICIQ-OABqol), and a participant satisfaction questionnaire. Therapy responders were defined as participants with ≥50% reduction in UUI episodes compared to baseline. Data were analyzed on all 129 participants.
Results: At 1 year, 89% of the participants were therapy responders. The average UUI episodes per day reduced from 5.6 ± 0.3 at baseline to 1.4 ± 0.2. Participants experienced an overall clinically meaningful improvement of 34 points on the ICIQ-OABqol questionnaire. All study participants (100%) were able to recharge their device at 1 year, and 96% of participants reported that the frequency and duration of recharging was acceptable. There were no serious device-related adverse events.
Conclusions: The Axonics System is safe and effective at 1 year, with 89% of participants experiencing clinically and statistically significant improvements in UUI symptoms.
Keywords: clinical trial; implantable neurostimulator; overactive bladder; sacral neuromodulation; urinary urgency incontinence.
© 2020 The Authors. Neurourology and Urodynamics Published by Wiley Periodicals, Inc.