Background: Noncompressible torso hemorrhage remains a leading cause of death. Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA) placement may occur before transport; however, its efficacy has not been demonstrated at altitude. We hypothesized that changes in altitude would not result in blood pressure changes proximal to a deployed REBOA.
Methods: A simulation model for 7Fr guidewireless REBOA was used at altitudes up to 22,000 feet. Female pigs then underwent hemorrhagic shock to a mean arterial pressure (MAP) of 40 mm Hg. After hemorrhage, a REBOA catheter was deployed in the REBOA group and positioned but not inflated in the no-REBOA group. Animals underwent simulated aeromedical evacuation at 8000 ft or were left at ground level. After altitude exposure, the balloon was deflated, and the animals were observed.
Results: Taking the REBOA catheter to 22,000 ft in the simulation model resulted in a lower systolic blood pressure but a preserved MAP. In the porcine model, REBOA increased both systolic blood pressure and MAP compared with no-REBOA (P < 0.05) and was unaffected by altitude. No differences in postflight blood pressure, acidosis, or systemic inflammatory response were observed between ground and altitude REBOA groups.
Conclusions: REBOA maintained MAP up to 22,000 feet in an inanimate model. In the porcine model, REBOA deployment improved MAP, and the balloon remained effective at altitude.
Keywords: Aeromedical evacuation; Altitude resuscitation; REBOA; Tactical flight; Trauma.