Activity measurements of 3H, 241Am and 60Co solutions were performed to compare digital coincidence modules used at PTB and POLATOM for TDCR and 4πβ(LS)-γ coincidence counting. The activities determined with various coincidence modules connected in parallel to the same counter at PTB were found to be consistent. Observed discrepancies caused by differences in the coincidence resolving time did not exceed 0.14%. Accidental coincidences simulated by a frequency generator were registered, and the coincidence resolving time was determined.
Keywords: 4πβ(LS)-γ coincidence system; Digital coincidence module; TDCR.