Purpose: To assess the use of image registration for correcting respiratory motion in free breathing lung T1 mapping acquisition in patients with idiopathic pulmonary fibrosis (IPF).
Theory and methods: The method presented used image registration to synthetic images during postprocessing to remove respiratory motion. Synthetic images were generated from a model of the inversion recovery signal of the acquired images that incorporated a periodic lung motion model. Ten healthy volunteers and 19 patients with IPF underwent 2D Look-Locker T1 mapping acquisition at 1.5T during inspiratory breath-hold and free breathing. Eight healthy volunteers and seven patients with IPF underwent T1 mapping acquisition during expiratory breath-hold. Fourteen patients had follow-up scanning at 6 months. Dice similarity coefficient (DSC) was used to evaluate registration efficacy.
Results: Image registration increased image DSC (P < .001) in the free breathing inversion recovery images. Lung T1 measured during a free breathing acquisition was lower in patients with IPF when compared with healthy controls (inspiration: P = .238; expiration: P = .261; free breathing: P = .021). Measured lung T1 was higher in expiration breath-hold than inspiration breath-hold in healthy volunteers (P < .001) but not in patients with IPF (P = .645). There were no other significant differences between lung T1 values within subject groups.
Conclusions: The registration technique significantly reduced motion in the Look-Locker images acquired during free breathing and may improve the robustness of lung T1 mapping in patients who struggle to hold their breath. Lung T1 measured during a free breathing acquisition was significantly lower in patients with IPF when compared with healthy controls.
Keywords: Lung; T1 map; diffuse lung disease; image registration; motion correction.
© 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.