Introduction: Good osteointegration of implants requires sufficient primary stability. Aim of this study was to examine the influence of screw fixation on the primary stability of press-fit cups.
Methods: Two press-fit acetabular cups were tested with regard to the influence and number of screws and their primary stability.
Results: For the relatively thin-walled Allofit®-S cup, an influence of the number of screws on the different forms of movement could be demonstrated.
Conclusions: We see a clear influence of the cup wall thickness on the elastic deformability and accordingly on the primary stability of the examined cups.
Keywords: Cementless; Cup component; Hip endoprosthetics; Primary stability; Screw connection.