Scintillation imaging as a high-resolution, remote, versatile 2D detection system for MR-linac quality assurance

Med Phys. 2020 Sep;47(9):3861-3869. doi: 10.1002/mp.14353. Epub 2020 Jul 18.

Abstract

Purpose: To demonstrate the potential benefits of remote camera-based scintillation imaging for routine quality assurance (QA) measurements for magnetic resonance guided radiotherapy (MRgRT) linear accelerators.

Methods: A wall-mounted CMOS camera with a time-synchronized intensifier was used to image photons produced from a scintillation screen in response to dose deposition from a 6 MV FFF x-ray beam produced by a 0.35 T MR-linac. The oblique angle of the field of view was corrected using a projective transform from a checkerboard calibration target. Output sensitivity and constancy was measured using the scintillator and benchmarked against an A28 ion chamber. Field cross-plane and in-plane profiles were measured for field sizes ranging from 1.68 × 1.66 cm2 to 20.02 × 19.92 cm2 with both scintillation imaging and using an IC profiler. Multileaf collimator (MLC) shifts were introduced to test sensitivity of the scintillation imaging system to small spatial deviations. A picket fence test and star-shot were delivered to both the scintillator and EBT3 film to compare accuracy in measuring MLC positions and isocenter size.

Results: The scintillation imaging system showed comparable sensitivity and linearity to the ion chamber in response to changes in machine output down to 0.5 MU (R2 = 0.99). Cross-plane profiles show strong agreement with defined field sizes using full width half maximum (FWHM) measurement of <2 mm for field sizes below 15 cm, but the oblique viewing angle was the limiting factor in accuracy of in-plane profile widths. However, the system provided high-resolution profiles in both directions for constancy measurements. Small shifts in the field position down to 0.5 mm were detectable with <0.1 mm accuracy. Multileaf collimator positions as measured with both scintillation imaging and EBT3 film were measured within ± 1 mm tolerance and both detection systems produced similar isocenter sizes from the star-shot analysis (0.81 and 0.83 mm radii).

Conclusions: Remote scintillation imaging of a two-dimensional screen provided a rapid, versatile, MR-compatible solution to many routine quality assurance procedures including output constancy, profile flatness and symmetry constancy, MLC position verification and isocenter size. This method is high-resolution, does not require post-irradiation readout, and provides simple, instantaneous data acquisition. Full automation of the readout and processing could make this a very simple but effective QA tool, and is adaptable to all medical accelerators.

Keywords: MR-linac; QA; optical imaging; scintillation.

MeSH terms

  • Calibration
  • Particle Accelerators*
  • Photons
  • Radiotherapy, Image-Guided*