Objective: Stentless aortic valves have shown superior hemodynamic performance and faster left ventricular mass regression compared to stented bioprostheses. Yet, controversies exist concerning the durability of stentless valves. This case-matched study compared short- and long-term clinical outcomes of stentless LivaNova-Sorin Pericarbon Freedom™ (SPF) and stented Carpentier-Edwards Perimount (CEP) aortic prostheses.
Methods: From 2003 through 2006, 134 consecutive patients received aortic valve replacement with SPF at our institution. This cohort was matched, according to 20 preoperative clinical parameters, with a control group of 390 patients who received CEP prosthesis during the same time. The resulting 55 + 55 matched patients were analyzed for perioperative results and long-term clinical outcomes.
Results: Early mortality was 0% for both groups. Lower transvalvular gradients were found in the SPF group (10.6 ± 2.9 versus 15.7 ± 3.1 mmHg, P < 0.001). Overall late mortality (mean follow-up: 10.03 years) was similar for both groups (50.1% versus 42.8%, P = 0.96). Freedom from structural valve degeneration (SVD) at 13 years was similar for both groups (SPF = 92.3%, CEP = 73.9%, P = 0.06). Freedom from aortic valve reinterventions did not differ (SPF = 92.3%, CEP = 93.5%, P = 0.55). Gradients at 13-year follow-up remained significantly lower in SPF group (10.0 ± 4.5 versus 16.2 ± 9.5 mmHg, P < 0.001). Incidence of acute bacterial endocarditis (ABE) and major adverse cardiovascular and cerebrovascular events (MACCE) was similar.
Conclusions: SPF and CEP demonstrated comparable long-term outcomes related to late mortality, SVD, aortic valve reinterventions, and incidence of ABE and MACCE. Superior hemodynamic performance of SPF over time can make this valve a suitable choice in patients with small aortic root and large body surface area.
Keywords: aortic valve replacement; propensity matching; stented valve; stentless valve.