We report the case of a bronchial foreign body, following a tracheostomy site swab for SARS-CoV-2, aiming to raise awareness and vigilance. A qualified nurse was performing a routine SARS-CoV-2 swab on a 51-year-old woman, fitted with a tracheostomy in the recent past following a craniotomy. This was part of the discharging protocol to a nursing home. During the sampling, part of the swab stylet snapped and was inadvertently dropped through the tracheostomy site. Initial CT imaging was reported as showing no signs of a foreign body but some inflammatory changes. Bedside flexible endoscopy through the tracheostomy site revealed the swab in a right lobar bronchus. This was subsequently removed by flexible bronchoscopy. This case highlights the need for clear guidance on how samples for SARS-CoV-2 are taken from patients with front of neck airways (laryngectomy/tracheοstomy) and the potential pitfalls involved.
Keywords: ear; medical-surgical nursing; nose and throat/otolaryngology; otolaryngology / ENT; radiology.
© BMJ Publishing Group Limited 2020. No commercial re-use. See rights and permissions. Published by BMJ.