Objectives: The Florida-sleeve is a valve-sparing technique that causes minimal interference to leaflet kinematics and aortic root dynamism. The aim of this in vitro study was to evaluate the effects of the Florida-sleeve and Yacoub techniques on aortic leaflet kinematics.
Methods: Two groups of 6 whole porcine hearts were treated with either the Florida-sleeve technique or the Yacoub technique and tested in a pulsatile loop. Valve fluid dynamics, coronary flow analysis and valve echocardiograms were performed both before and after the procedures.
Results: Both procedures showed no difference in rapid valve opening time as compared with their respective baseline values. The Florida-sleeve procedure showed a shorter slow closing time (192 ± 19 ms vs baseline 244 ± 14 ms, P = 0.016) and increased slow closing velocity (-1.5 ± 0.4 cm/s vs baseline -0.8 ± 0.4 cm/s, P = 0.038). In the rapid valve closing phase, the Yacoub procedure showed a trend towards slower closing valve velocity (-16 ± 9 cm/s vs baseline -25 ± 9 cm/s, P = 0.07). The Yacoub procedure showed larger leaflet displacement at the end of the slow valve closing time that was 2.0 ± 0.5 cm vs baseline 1.5 ± 0.3 cm, P = 0.044. When comparing the Florida-sleeve and Yacoub procedures, the former showed statistically significant shorter slow valve closing time (P = 0.017).
Conclusions: This study showed that the Florida-sleeve technique alters the slow closing phase of the aortic valve leaflet kinematics when compared with both the normal baseline and Yacoub procedure, while the latter showed a larger leaflet displacement before the rapid closing valve phase.
Keywords: Adult; Aortic valve sparing; Leaflet kinematics.
© The Author(s) 2020. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.