Diagnostic accuracy of computer aided electrocardiogram analysis in dogs

J Small Anim Pract. 2021 Feb;62(2):145-149. doi: 10.1111/jsap.13267. Epub 2020 Dec 1.

Abstract

Objectives: Evaluation of a computerised electrocardiogram algorithm compared to the interpretation of a team of board-certified veterinary cardiologists.

Materials and methods: This was a cross-sectional retrospective cohort study. A total of 399 electronic canine electrocardiogram recordings screened from 1391 electrocardiograms were enrolled in the study. A panel of seven cardiologists, masked to patient information, evaluated electrocardiograms for the following: P-wave amplitude and duration; PR-interval; R-wave amplitude; QRS duration; heart rate; mean electrical axis; and final overall diagnosis for the detection of arrhythmia and any abnormal electrocardiogram anomaly.

Results: The sensitivity of the electrocardiogram algorithm for detecting arrhythmias was 99.7% (95% confidence intervals, CI: 98.5 to 99.9) and the specificity was 99.5% (95% CI: 98.0 to 99.9) compared to the consensus result created by panel of cardiologists. The sensitivity of the algorithm for the detection of any electrocardiogram anomaly, including abnormal measurements, was 71.3% (95% CI: 65.5 to 76.7) and the specificity was 35.1% (95% CI: 27.0 to 43.8) compared to the panel of cardiologists.

Clinical significance: The electrocardiogram algorithm was shown to have high sensitivity for the detection of arrhythmias, but not all electrocardiogram anomalies. The results support the use of this algorithm as a tool to aid in the triage of the electrocardiogram workflow.

MeSH terms

  • Animals
  • Arrhythmias, Cardiac / diagnosis
  • Arrhythmias, Cardiac / veterinary
  • Computers
  • Cross-Sectional Studies
  • Dog Diseases* / diagnosis
  • Dogs
  • Electrocardiography* / veterinary
  • Retrospective Studies