Introduction: This study aimed to assess and compare the deformation that develops in the circummaxillary sutures during activation of micro-implant assisted rapid palatal expander (MARPE) and tooth-supported expander (Hyrax) s, in the rapid maxillary expansion.
Settings and sample: 7 pigs Sus Scrofa received custom-made MARPE (n = 3) and Hyrax (n = 4) appliances.
Material and methods: The devices were activated 25 times with strain readings captured by strain gauges attached to the following regions: posterior midpalatal suture (MPS), maxilla-premaxilla suture (MPM), maxilla-zygomatic suture (MZ) and maxilla pterygoid-process suture (MPP). The intermolar distance and suture width were measured immediately before activation and at the 20th and 25th activation. ANOVA and Kruskal-Wallis test was applied.
Results: The MARPE group presented greater MPS displacement in all measured regions, and one of the devices produced a significant opening (1.7 mm) in the posterior region. The accumulated tension in the MPS was higher compared to the other sutures (P < .05). A MARPE animal presented higher median tension in the MPS region (294.77με) compared to all other animals except one Hyrax animal (P < .05). Regarding the median tensions of the different activation intervals, the median tension measured during the 16th to 25th activation interval in the Hyrax group was lower than that measured during the first 8 activations, in both the MPS and MZ (P < .05).
Conclusions: MARPE expanders developed more constant tensions during all activations (MPS and MZ), while Hyrax showed lower tension in the 16th to 25th activation.
Keywords: equipment and supplies; orthodontic anchorage procedures; palatal expansion technique.