Comparison of ELISA- and SIMOA-based quantification of plasma Aβ ratios for early detection of cerebral amyloidosis

Alzheimers Res Ther. 2020 Dec 5;12(1):162. doi: 10.1186/s13195-020-00728-w.

Abstract

Background: Blood-based amyloid biomarkers may provide a non-invasive, cost-effective and scalable manner for detecting cerebral amyloidosis in early disease stages.

Methods: In this prospective cross-sectional study, we quantified plasma Aβ1-42/Aβ1-40 ratios with both routinely available ELISAs and novel SIMOA Amyblood assays, and provided a head-to-head comparison of their performances to detect cerebral amyloidosis in a nondemented elderly cohort (n = 199). Participants were stratified according to amyloid-PET status, and the performance of plasma Aβ1-42/Aβ1-40 to detect cerebral amyloidosis was assessed using receiver operating characteristic analysis. We additionally investigated the correlations of plasma Aβ ratios with amyloid-PET and CSF Alzheimer's disease biomarkers, as well as platform agreement using Passing-Bablok regression and Bland-Altman analysis for both Aβ isoforms.

Results: ELISA and SIMOA plasma Aβ1-42/Aβ1-40 detected cerebral amyloidosis with identical accuracy (ELISA: area under curve (AUC) 0.78, 95% CI 0.72-0.84; SIMOA: AUC 0.79, 95% CI 0.73-0.85), and both increased the performance of a basic demographic model including only age and APOE-ε4 genotype (p ≤ 0.02). ELISA and SIMOA had positive predictive values of respectively 41% and 36% in cognitively normal elderly and negative predictive values all exceeding 88%. Plasma Aβ1-42/Aβ1-40 correlated similarly with amyloid-PET for both platforms (Spearman ρ = - 0.32, p < 0.0001), yet correlations with CSF Aβ1-42/t-tau were stronger for ELISA (ρ = 0.41, p = 0.002) than for SIMOA (ρ = 0.29, p = 0.03). Plasma Aβ levels demonstrated poor agreement between ELISA and SIMOA with concentrations of both Aβ1-42 and Aβ1-40 measured by SIMOA consistently underestimating those measured by ELISA.

Conclusions: ELISA and SIMOA demonstrated equivalent performances in detecting cerebral amyloidosis through plasma Aβ1-42/Aβ1-40, both with high negative predictive values, making them equally suitable non-invasive prescreening tools for clinical trials by reducing the number of necessary PET scans for clinical trial recruitment.

Trial registration: EudraCT 2009-014475-45 (registered on 23 Sept 2009) and EudraCT 2013-004671-12 (registered on 20 May 2014, https://www.clinicaltrialsregister.eu/ctr-search/trial/2013-004671-12/BE ).

Keywords: Biomarkers; Cerebral amyloidosis; ELISA; Immunoassay; Plasma; Preclinical Alzheimer’s disease; Prescreening; SIMOA; β-Amyloid.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Alzheimer Disease*
  • Amyloid beta-Peptides
  • Amyloidosis* / diagnostic imaging
  • Biomarkers
  • Cross-Sectional Studies
  • Enzyme-Linked Immunosorbent Assay
  • Humans
  • Peptide Fragments
  • Prospective Studies

Substances

  • Amyloid beta-Peptides
  • Biomarkers
  • Peptide Fragments