Improved Ultraviolet Radiation Film Dosimetry Using OrthoChromic OC-1 Film

Photochem Photobiol. 2021 May;97(3):498-504. doi: 10.1111/php.13364. Epub 2020 Dec 28.

Abstract

There is growing interest in far-UVC lighting, defined as wavelengths from 200 to 230 nm, because research has demonstrated these wavelengths to be an effective antimicrobial technology while posing a minimal hazard to human health. Far-UVC lighting is now being installed to directly irradiate spaces where humans are present, and it will be important to perform measurements to verify far-UVC lighting installations are operating within widely accepted exposure guidelines. In this work, we explore the use of a commercially available film, known as OrthoChromic OC-1, to measure ultraviolet radiation exposure. The film was tested with a variety of ultraviolet wavelengths and irradiance conditions, and the color change of the film was analyzed for increasing levels of radiant exposure. The film response extended over a dynamic range that was greater than the recommended exposure limits for far-UVC radiation so it can potentially be useful for health hazard monitoring. The spectrum of the incident ultraviolet radiation strongly affected the response of the film; therefore, for accurate measurements we recommend the measured spectrum match the spectrum used for calibration. Overall, dosimetry with this film provides a simple, accurate, and inexpensive method of quantifying ultraviolet radiation exposure that is suitable for far-UVC measurements.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Calibration
  • Film Dosimetry*
  • Humans
  • Lighting
  • Ultraviolet Rays*