Dosimetric assessment of brass mesh bolus and transparent polymer-gel type bolus for commonly used breast treatment delivery techniques

Med Dosim. 2021;46(3):e10-e14. doi: 10.1016/j.meddos.2021.01.001. Epub 2021 Jan 31.

Abstract

We investigated skin dose enhancements of brass mesh bolus (BMB) and a recently developed transparent polymer-gel bolus (PGB) for clinically relevant breast treatment delivery techniques. The dose enhancement of the breast surface with BMB and PGB were compared to that of tissue-equivalent bolus. Three breast treatment plans were generated on CT scans of an anthropomorphic chest phantom: tangential step-and-shoot 3D conformal (3DCRT) planned using Field-in-Field (FiF), tangential sliding-window 3DCRT using Electronic Compensator (EC), and volumetric modulated arc therapy (VMAT). All plans were created using 6 MV photons and a prescription dose (Rx) of 180 cGy per fraction. Skin doses of all 3 plans were measured with radiochromic films, separately delivered in triplicate. Each plan was delivered to the phantom without bolus, and then with BMB (1 or 2 layers; 3 or 10 mm tissue-equivalent), PGB, and Superflab (3, 5, and 10 mm tissue-equivalent). Doses were determined by reading the radiochromic films with a flatbed scanner, and analyzing the images using a calibration curve for each specific batch. For all bolus types and plans, surface doses averaged over the 3 measurements were between 88.4% and 107.4% of Rx. Without bolus, average measured skin doses were between 51.2% and 64.2% of Rx. Skin doses with BMB and PGB were comparable to that with tissue-equivalent bolus. Over all 3 treatment delivery techniques, using BMB resulted in average skin doses of 92.8% and 102.1% for 1- and 2 layers, respectively, and using PGB results in average skin doses of 94.8%, 98.2%, and 99.7% for 3, 5, and 10-mm tissue-equivalent, respectively. The average measured skin doses with BMB and PGB agreed within ± 3% compared to the tissue-equivalent thickness bolus. We concluded that BMB and PGB are clinically equivalent in skin dose enhancement for breast treatment as the 3, 5, and 10 mm tissue-equivalent bolus.

Keywords: Bolus; Breast treatment; Dose measurement; Skin dose.

MeSH terms

  • Copper
  • Humans
  • Polymers
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted*
  • Radiotherapy, Intensity-Modulated*
  • Surgical Mesh
  • Zinc

Substances

  • Polymers
  • brass
  • Copper
  • Zinc