New image quality and dose reduction technique for pediatric digital radiography

Diagn Interv Imaging. 2021 Jul-Aug;102(7-8):463-470. doi: 10.1016/j.diii.2021.01.009. Epub 2021 Feb 3.

Abstract

Purpose: The purpose of this study was to test a new post-processing and denoising engine for patient dose reduction while maintaining diagnostic image quality (IQ) in pediatric digital radiography (DR).

Materials and methods: Pediatric DR images of the thorax, pelvis, abdomen and spine obtained in 174 patients (102 males, 72 females; mean age, 2±1.8 [SD] years; age range: 6 months-9 years) were retrieved. Artificial noise was added to the images to simulate acquisitions at 50%, 32% and 12.5% of the routine dose levels. A total of 696 images corresponding to four dose levels were post-processed using S-Vue™ and further blindly scored by three pediatric radiologists using a scoring grid of 4-6 criteria specifically defined per anatomical area. The mean score was assessed for each area and weight class (5-15 and 15-30kg) and compared across the simulated low dose images. Paired Wilcoxon test was used with a threshold difference of 0.5 (half a criterion) between scores to highlight a significant reduction in image quality. Inter-rater reliability was assessed using intraclass correlation coefficient (ICC).

Results: Only the 50% reduced dose images showed non-inferiority when compared to routine images for all of areas and weight classes (P<0.01). Very good inter-rater reliability of the overall scores was observed for the pelvis in the 5-15kg weight class (ICC=0.85) for images at full dose, 50% and 32% reduced dose. For the remaining areas (thorax, abdomen and spine) and weight classes, inter-rater reliability was moderate (ICC: 0.3-0.6).

Conclusion: S-Vue™ post-processing software allows a two-fold radiation dose reduction while maintaining satisfactory IQ in pediatric DR.

Keywords: Digital radiography; Dose reduction; Image noise; Image processing; Pediatrics.