Aims: Transvenous lead extraction is associated with a significant risk of complications and identifying patients at highest risk pre-procedurally will enable interventions to be planned accordingly. We developed the ELECTRa Registry Outcome Score (EROS) and applied it to the ELECTRa registry to determine if it could appropriately risk-stratify patients.
Methods and results: EROS was devised to risk-stratify patients into low risk (EROS 1), intermediate risk (EROS 2), and high risk (EROS 3). This was applied to the ESC EORP European Lead Extraction ConTRolled ELECTRa registry; 57.5% EROS 1, 31.8% EROS 2, and 10.7% EROS 3. Patients with EROS 3 or 2 were significantly more likely to require powered sheaths and a femoral approach to complete procedures. Patients with EROS 3 were more likely to suffer procedure-related major complications including deaths (5.1 vs. 1.3%; P < 0.0001), both intra-procedural (3.5 vs. 0.8%; P = 0.0001) and post-procedural (1.6 vs. 0.5%; P = 0.0192). They were more likely to suffer post-procedural deaths (0.8 vs. 0.2%; P 0.0449), cardiac avulsion or tear (3.8 vs. 0.5%; P < 0.0001), and cardiovascular lesions requiring pericardiocentesis, chest tube, or surgical repair (4.6 vs. 1.0%; P < 0.0001). EROS 3 was associated with procedure-related major complications including deaths [odds ratio (OR) 3.333, 95% confidence interval (CI) 1.879-5.914; P < 0.0001] and all-cause in-hospital major complications including deaths (OR 2.339, 95% CI 1.439-3.803; P = 0.0006).
Conclusion: EROS successfully identified patients who were at increased risk of significant procedural complications that require urgent surgical intervention.
Keywords: ELECTRa Registry Outcome Score; ELECTRa registry; Risk stratification; Transvenous lead extraction.
Published on behalf of the European Society of Cardiology. All rights reserved. © The Author(s) 2021. For permissions, please email: [email protected].