Impaired performance of SARS-CoV-2 antigen-detecting rapid diagnostic tests at elevated and low temperatures

J Clin Virol. 2021 May:138:104796. doi: 10.1016/j.jcv.2021.104796. Epub 2021 Mar 16.

Abstract

Antigen-detecting rapid diagnostic tests (Ag-RDTs) can complement molecular diagnostics for COVID-19. The recommended temperature for storage of SARS-CoV-2 Ag-RDTs ranges between 2-30 °C. In the global South, mean temperatures can exceed 30 °C. In the global North, Ag-RDTs are often used in external testing facilities at low ambient temperatures. We assessed analytical sensitivity and specificity of eleven commercially-available SARS-CoV-2 Ag-RDTs using different storage and operational temperatures, including short- or long-term storage and operation at recommended temperatures or at either 2-4 °C or at 37 °C. The limits of detection of SARS-CoV-2 Ag-RDTs under recommended conditions ranged from 1.0×106- 5.5×107 genome copies/mL of infectious SARS-CoV-2 cell culture supernatant. Despite long-term storage at recommended conditions, 10 min pre-incubation of Ag-RDTs and testing at 37 °C resulted in about ten-fold reduced sensitivity for five out of 11 SARS-CoV-2 Ag-RDTs, including both Ag-RDTs currently listed for emergency use by the World Health Organization. After 3 weeks of storage at 37 °C, eight of the 11 SARS-CoV-2 Ag-RDTs exhibited about ten-fold reduced sensitivity. Specificity of SARS-CoV-2 Ag-RDTs using cell culture supernatant from common respiratory viruses was not affected by storage and testing at 37 °C, whereas false-positive results occurred at outside temperatures of 2-4 °C for two out of six tested Ag-RDTs, again including an Ag-RDT recommended by the WHO. In summary, elevated temperatures impair sensitivity, whereas low temperatures impair specificity of SARS-CoV-2 Ag-RDTs. Consequences may include false-negative test results at clinically relevant virus concentrations compatible with transmission and false-positive results entailing unwarranted quarantine assignments. Storage and operation of SARS-CoV-2 Ag-RDTs at recommended conditions is essential for successful usage during the pandemic.

Keywords: Rapid antigen test; SARS-CoV-2; Sensitivity; Specificity; Temperature stability; Tropics; Winter.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • COVID-19 / diagnosis*
  • COVID-19 Serological Testing*
  • Cold Temperature / adverse effects
  • Diagnostic Tests, Routine*
  • False Negative Reactions
  • False Positive Reactions
  • Hot Temperature / adverse effects
  • Humans
  • Reagent Kits, Diagnostic*
  • Sensitivity and Specificity