Validation of the Thermo Scientific™ SARS-CoV-2 RT-PCR Detection Workflow for the Detection of SARS-CoV-2 from Stainless-Steel Environmental Surface Swabs: AOAC Performance Tested MethodSM 012103

J AOAC Int. 2021 Aug 20;104(4):935-947. doi: 10.1093/jaoacint/qsab050.

Abstract

Background: The Thermo Scientific™ SARS-CoV-2 reverse transcription-polymerase chain reaction (RT-PCR) Detection Workflow, packaged with Applied Biosystems™ TaqMan™ 2019-nCoV Assay Kit v1 targets three different SARS-CoV-2 genomic regions in a single RT-PCR reaction.

Objective: To validate the Thermo Scientific SARS-CoV-2 RT-PCR Workflow, for the detection of SARS-CoV-2 virus on stainless-steel surfaces as part of the AOAC Performance Tested MethodSM Emergency Response Validation program.

Method: The Applied Biosystems TaqMan 2019-nCoV Assay Kit v1, as part of the Thermo Scientific SARS-CoV-2 RT-PCR Workflow, was evaluated for specificity using in silico analysis of 15 764 SARS-CoV-2 sequences and 65 exclusivity organisms. The Thermo Scientific SARS-CoV-2 RT-PCR Workflow was evaluated in an unpaired study for one environmental surface (stainless steel) and compared to the U.S. Centers for Disease Control and Prevention 2019-Novel Coronavirus RT-PCR Diagnostic Panel, Instructions for Use (Revision 4, Effective 6/12/2020).

Results: In silico analysis showed that, of the 15 756 target SARS-CoV-2 genomes analyzed, 99% of the strains/isolates are perfectly matched to at least two of the three assays, and more than 90% have 100% homology to all three assays (ORF1ab, N-gene, S-gene) in the SARS-CoV-2 Kit. None of the 65 non-target strain genomes analyzed showed matching sequences. In the matrix study, the Thermo Scientific SARS-CoV-2 workflow showed comparable detection to the centers of disease control and prevention (CDC) method.

Conclusions: The Thermo Scientific SARS-CoV-2 RT-PCR Workflow is an effective procedure for detection of RNA from SARS-CoV-2 virus from stainless steel.

Highlights: The workflow provides equivalent performance results with the two tested RNA extraction platforms and the two tested RT-PCR instruments.

MeSH terms

  • COVID-19*
  • Humans
  • RNA, Viral / genetics
  • Reverse Transcriptase Polymerase Chain Reaction
  • SARS-CoV-2*
  • Sensitivity and Specificity
  • Stainless Steel
  • Workflow

Substances

  • RNA, Viral
  • Stainless Steel