Outcomes of Prosthesis-Patient Mismatch Following Supra-Annular Transcatheter Aortic Valve Replacement: From the STS/ACC TVT Registry

JACC Cardiovasc Interv. 2021 May 10;14(9):964-976. doi: 10.1016/j.jcin.2021.03.040.

Abstract

Objectives: The aim of this study was to assess the outcomes of severe prosthesis-patient mismatch (PPM) in the TVT (Transcatheter Valve Therapy) Registry in patients undergoing supra-annular transcatheter aortic valve replacement (TAVR) for de novo stenosis or failed surgical bioprostheses (transcatheter aortic valve [TAV]-in-surgical aortic valve [SAV]).

Background: Severe PPM has been associated with adverse outcomes following TAVR, yet the clinical outcome of severe PPM after supra-annular TAVR is largely unknown.

Methods: Supra-annular TAVR was performed in patients enrolled in the TVT Registry with de novo stenosis (n = 42,174) or TAV-in-SAV (n = 5,446). Valve Academic Research Consortium-3 criteria were used to define severe PPM. The clinical impact of severe PPM on 1-year mortality and valve-related readmission was assessed using multivariate regression. A generalized linear mixed model was used to evaluate predictors of severe PPM.

Results: Severe PPM was found in 5.3% of patients undergoing de novo TAVR and 27.0% of patients undergoing TAV-in-SAV. The presence of severe PPM was not significantly associated with 1-year mortality or valve-related readmissions in both groups. Mean aortic gradients were higher in patients with severe PPM than in those without severe PPM at 1 month (9.7 ± 5.7 mm Hg vs. 7.3 ± 4.0 mm Hg; p < 0.001) and 1 year (10.2 ± 6.4 mm Hg vs. 8.0 ± 4.3 mm Hg; p < 0.001). Pre-procedural factors, including a <20-mm aortic annulus, were positive predictors of severe PPM in patients undergoing de novo TAVR (area under the curve = 0.795) and TAV-in-SAV (area under the curve = 0.764).

Conclusions: Severe PPM after supra-annular TAVR was not associated with increased 1-year mortality or valve-related readmissions. Longer-term follow-up is needed to determine if higher residual gradients in patients with severe PPM predict long-term outcomes. (STS/ACC Transcatheter Valve Therapy Registry [TVT Registry]; NCT01737528).

Keywords: aortic valve stenosis; patient-prosthesis mismatch; supra-annular; transcatheter aortic valve replacement.

Publication types

  • Clinical Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Heart Valve Prosthesis*
  • Humans
  • Prosthesis Design
  • Registries
  • Risk Factors
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01737528