Enabling LIFDI-MS measurements of highly air sensitive organometallic compounds: a combined MS/glovebox technique

Dalton Trans. 2021 Jul 6;50(26):9031-9036. doi: 10.1039/d1dt00978h.

Abstract

A new setup combining a ThermoFisher Exactive Plus Orbitrap Mass Spectrometer with a liquid injection field desorption ionization (LIFDI) source directly connected to an inert atmosphere glovebox is presented. The described setup allows for the analysis of very air- and moisture sensitive samples. Furthermore, the soft nature of LIFDI ionization gives access to the molecular ions of fragile molecules. This new setup is therefore especially useful for sensitive organometallic complexes. The functionality of the new setup is tested against [(Cp)2TiCl]˙, which is known for its notorious sensitivity to air and moisture. Its drastic colour change from green to orange upon exposure to air further supports the easy detection of traces of oxygen during the experiment. In addition, we applied this setup to the mass spectrometric analysis of the qualitative composition of a Cu/Al cluster mixture, which is not accessible by other analytical methods.