In this work, a novel, to the best of our knowledge, approach based on an x-ray thin lens imaging theory is proposed to predict the angular sensitivity responses of dual-phase-grating differential phase contrast (DPC) interferometers. Experimental validations have been performed to demonstrate the high accuracy of theoretical predictions using two different setups: one with real source images and the other with virtual source images. This new sensitivity calculation method is helpful to optimize the DPC imaging performance of a dual-phase-grating system.