Increasing the height of the anterior lip on a tibial insert in a posterior stabilized knee prosthesis has little effect on the wear rate

Med Eng Phys. 2021 May:91:48-53. doi: 10.1016/j.medengphy.2021.03.007. Epub 2021 Mar 27.

Abstract

A high anterior lip on a total knee prosthesis is an effective way of reducing anterior translation, but the effect on joint wear is unclear. Using finite element analysis (FEA), this study quantitatively compared wear rates and anterior contact stresses in three posterior stabilized knee prostheses with different heights for the anterior lip during six daily activities (walking, stair ascent, stair descent, sit-to-stand, pivot turn and crossover turn). The wear rate and location of maximum wear depth were similar for the three lip heights tested, but the knee with the highest anterior lip also showed slight anterior wear scaring due to articular contact stress during swing phase, which was highly dependent on the shape of the contact interface. This study illustrates that tibial inserts with a high anterior lip maintain a wear rate similar to moderate and low lip posterior stabilized designs.

Keywords: Contact stresses; Finite element method; High anterior lip; Knee prosthesis; Wear.