Haemolysis in prehospital blood samples

Scand J Clin Lab Invest. 2021 Oct;81(6):508-510. doi: 10.1080/00365513.2021.1941234. Epub 2021 Jun 28.

Abstract

The increasing use of Point Of Care Testing (POCT) in the prehospital setting demands a high and consistent quality of blood samples. We have investigated the degree of haemolysis in 779 prehospital blood samples and found a significant increase in haemolysis compared to intrahospital samples. The degree of haemolysis was within acceptable limits for current analyses. However, haemolysis should be taken into account when implementing future analyses in the prehospital field.

Keywords: Hemolysis; blood specimen collection; emergency care; point-of-care testing; pre-analytical phase; prehospital.

MeSH terms

  • Aged
  • Blood Specimen Collection*
  • Hemolysis / physiology*
  • Hospitals*
  • Humans