Objective: Surgeons have no direct objective feedback on cochlear-implant electrode array (EA) positioning during insertion, yet optimal hearing outcomes are contingent on placing the EA as close as feasible to viable neural endings. This paper describes a system to non-invasively determine intracochlear positioning of an EA, without requiring any modifications to existing commercial EAs themselves.
Methods: Electrical impedance has been suggested as a way to measure EA proximity to the inner wall of the cochlea that houses auditory nerve endings-the modiolus. In this paper, we extend prior work and demonstrate for the first time the relationship between bipolar access resistance and proximity of the EA to the modiolus (E-M proximity). We also evaluate two methods for producing direct, real-time estimates of E-M proximity from bipolar impedance measurements.
Results: We show that bipolar access resistance is highly correlated with E-M proximity and can be approximately modeled by a power law function. This one dimensional model is shown to be capable of producing accurate real-time estimates of E-M proximity, but its simplicity also limits the potential for future improvement. To address this challenge, we propose a new prediction approach based on a recurrent neural network, which generated an overall prediction accuracy of 93.7%.
Conclusion: Bipolar access resistance is highly correlated with E-M proximity, and can be used to estimate EA positioning.
Significance: This work shows how impedance sensing can be used to localize an EA during insertion into the small, enclosed cochlear environment, without requiring any modifications to existing clinically used EAs.