Background: Despite compliance to extensive reprocessing protocols, duodenoscopes have been linked to outbreaks of susceptible and multi-drug resistant organisms (MDRO) due to persistent duodenoscope contamination. Duodenoscope-associated infections (DAIs) based on transmission of susceptible microorganisms are likely to be underreported due to detection bias.
Case presentation: We describe the retrospective detection of a DAI case caused by a susceptible microorganism which at the time of clinical infection was not recognized as such. During 2017 and 2018, duodenoscopes were cultured on a daily basis due to research activities. While analyzing this data, it was found that a duodenoscope had been contaminated with Enterobacter cloacae complex over a period of 3 months. We checked whether patients treated with this duodenoscope had developed infections and found one patient with an E. cloacae cholangitis 3 months after the ERCP (Endoscopic retrograde cholangiopancreaticography) procedure. The isolates on the duodenoscope and in the patients' blood culture were indistinguishable by amplified fragment length polymorphism (AFLP). By classical multi-locus sequence typing (MLST), both strains were of the same (but novel) sequence type. Application of whole genome MLST showed 93 (out of 3757) allelic differences.
Conclusion: This case report describes a plausible link between a contaminated duodenoscope and a patient infection with E. cloacae. Transmission of susceptible E. cloacae was highly suspected from AFLP and MLST results; by WGS, 93 allelic differences were found which proves closely related strains. This report shows that DAIs by susceptible microorganisms can be easily missed and therefore its true prevalence remains underscored.
Keywords: Case report; Duodenoscope; Duodenoscope-associated infection (DAI); Endoscopic retrograde cholangiopancreaticography (ERCP); Nosocomial infection.
© 2021. The Author(s).