A New Electrochemiluminescence-Based Multiplex Assay for the Assessment of Human Antibody Responses to Bordetella pertussis Vaccines

Infect Dis Ther. 2021 Dec;10(4):2539-2561. doi: 10.1007/s40121-021-00530-7. Epub 2021 Sep 2.

Abstract

Introduction: Commercially available enzyme-linked immunosorbent assay (ELISA) kits designed for pertussis diagnostic purposes are frequently used to assess antibody responses to pertussis vaccines in clinical trials, but have limited accuracy and are not calibrated against international standards. We developed a new electrochemiluminescence (ECL)-based multiplexed assay and compared its performance to two commercial Bordetella pertussis ELISA kits and to historical in-house ELISAs.

Methods: The ECL assay quantifies serum concentrations of antibodies against four B. pertussis antigens: pertussis toxin (PT), filamentous hemagglutinin (FHA), pertactin (PRN), and fimbrial agglutinogen (FIM). The assay was validated for precision, accuracy, dilutability, lower limit of quantification, and specificity. Sera from a clinical trial (CTRI/2016/11/007434) were used to compare the ECL assay to two commercial ELISA kits available from GenWay BioTech and Demeditec Diagnostics for accuracy, linearity, specificity, and concordance to both internal (WWO-2-043) and international (NIBSC 06/140) references. Sera from four clinical trials (NCT02587520, NCT00255047, NCT00347958, NCT01346293) were used to compare the concordance to clinical ELISAs. Informed consent was ensured prior to using any sera.

Results: Precision, accuracy, dilutability, lower limit of quantification, and specificity were demonstrated for the ECL assay. Concordance between the ECL assay and established clinical ELISAs was met for antibody responses to PT, FIM, and PRN, but not for FHA. The ECL assay demonstrated higher accuracy and linearity than the ELISA kits. While concordance between the ECL and commercial kits was low, the ECL assay better distinguished between pre- and post-vaccination clinical samples.

Conclusion: The new ECL assay was validated for the quantitative evaluation of anti-PT, anti-FHA, anti-FIM, and anti-PRN IgG antibodies in samples from clinical trials, and demonstrated equivalent or better performance than two commercially available ELISA kits.

Keywords: Assay; Comparison; Development; ELISA; Electrochemiluminescence; Performance; Pertussis; Validation.