Background: The management of ureteral stricture is still a challenge for urologists. The aim of this prospective study was to assess the safety and effectiveness of self-expanding metal ureteral stents (URS) in ureteral strictures.
Methods: We performed URS placement procedures for ureteral stricture from Jan 2019 to July 2020, and prospectively collect various data before and after the operation. A paired T test was used to compare continuous variables before and after surgery, binary logistic regression analysis was used to identify the independent risk predictors of surgical failure.
Results: A total of 147 patients with 157 renal units received successful placement of URS. The mean operative time was 70.0 min. After a median follow-up time of 15 months, 73.2% (115/157) of stents were kept in situ. The most common complication was hematuria (13, 8.8%), followed by urinary tract infection (11, 7.5%) and pain (8, 5.4%). The volume of hydronephrosis (67.9 ± 34.9 VS 34.9 ± 51.1 cm3, P = 0.0001), serum creatinine level (103.0 ± 54.5 VS 93.8 ± 45.1 μmol/L, P = 0.034) and blood urea nitrogen level (6.6 ± 6.7 VS 5.4 ± 2.4 mmol/L, P = 0.032) decreased significantly at last follow up when compared with baseline. Stricture of the distal ureter was an independent risk factor for stent failure (HR 1.77, 95% CI 1.15, 2.73, P = 0.009).
Conclusions: URS was found to be safe and effective for ureteral strictures with a limited complications and good long-term results. For those who are not suitable for surgical reconstruction, the URS is an alternative management.
Keywords: Double-J stent; Metal stents; Ureteral stricture.